Figure/Table detail

Setup accuracy and dosimetric analysis of surface-guided radiotherapy in hypofractionated radiotherapy for breast cancer
ZHANG Weiheng, ZHAO Xuehui, ZOU Bing, LI Qing, ZHOU Zhenyu, LAN Qiongyu
INTERNATIONAL JOURNAL OF MEDICAL RADIOLOGY, 2026, 49(1): 50-57.   DOI: 10.19300/j.2026.L22196

组别 摆位误差/cm
X轴 Y轴 Z轴
常规组 0.08(-0.27,0.34) 0.06(-0.30,0.26) 0.06(-0.23,0.28)
SGRT组 0.06(-0.11,0.17) 0.05(-0.14,0.16) 0.03(-0.13,0.15)
W 3 059 4 254 4 120
P <0.001 <0.001 <0.001
表1 常规组和SGRT组摆位误差的比较 n=120